BNC 130 User Manual

Page 30

Advertising
background image

30

MAINTENANCE AND CALIBRATION


Pulse Dynamic Characteristics

Connect the 6040 TRIG OUT to the External Trigger input of the oscilloscope (bandwidth must be at
least 1 GHz). Connect the 6040 LIGHT OUT to the detector (also 1 GHz). Connect the detector's output
to channel A of the oscilloscope.

With the 6040 still in Pulse Mode, set the Peak level to 1.000 mW and the Baseline level to 100 µW. Set
TRIG for Internal Trigger, 100 kHz. Set Timing for a Width of 40 ns. and a Delay appropriate to view the
detector output on the scope. Verify that leading and trailing edges have transition times (10-90%) ≤ 1
ns.