4 replacing the inner door, 5 electrical components, Replacing the inner door – Eppendorf New Brunswick -86 °C Freezers Innova U360 User Manual

Page 33: Electrical components, Ma in ten ance

Advertising
background image

33

Innova® U360 -86 °C Freezers — Operating manual

3. Lift off inner door from hinges and set aside.

Abb. 13:

Lift-off inner door

Fig. 13:

Lift-off inner door

Repeat procedure for each door.

7.2.4

Replacing the inner door

1. Fully open the outer door of the freezer.
2. Fit door to hinge pins and close.
3. Check to ensure that inner door gasket is sealing against the freezer trim.
4. If required, adjust the latch retainer by loosening the screws and moving forward or

backwards.

5. Close outer door.

7.2.5

Electrical components

Audible alarms

Regularly check the audible alarm:

Press and hold the TEST key in the DIAGNOSTIC INFORMATION screen.

1 Inner door

2 Lift-off hinge

1

2

WARNING!

Risk of personal injury

All electrical components that could cause possible ignition of refrigerant vapor during normal
operation have been enclosed in an IP65 enclosure.

During routine maintenance, care must be taken to avoid any damage to the gaskets and
sealing grommets of these enclosures; also check the gaskets and sealing grommets
routinely to ensure their integrity. Should any damage or deformity be detected, the gasket
and/or sealing grommet must be replaced immediately.

Failure to observe this safety warning will invalidate the warranty and could result in a
dangerous situation.