Troubleshooting – Eppendorf New Brunswick -86 °C Freezers Innova U360 User Manual

Page 36

Advertising
background image

Innova® U360 -86 °C Freezers — Operating manual

36

Symptom/message

Cause

Remedy

HIGH CONDENSER
TEMP CHECK AIR
FILTER

Air-cooled condenser
temperature too high:
1. Filter may be blocked.

 
 

2. Ambient temperature may be

too high.

If alarm continues to sound:

Fan may have failed.

 
 

1. Clean filter according to the

instructions (see Air intake
grille and filter on p. 31)
.

2. Cool the room.
 

Call New Brunswick Service
department.

HIGH CONDENSER
TEMP CHECK ROOM
TEMP

Air-cooled condenser
temperature too high:
1. Ambient temperature may be

too high.

2. Freezer airflow may be

blocked or heated air from
other equipment is being
drawn into the freezer’s air
intake.

If alarm continues to sound:

Fan may have failed.

 
 

1. Reduce the room

temperature.

2. Remove restrictions to air flow

and make sure no other
equipment is forcing hot air
into the freezer intake.

 

Call New Brunswick Service
department.

EXCESSIVE
CONDENSER TEMP
THERMAL OVERLOAD
FREEZER WILL
REMAIN OFF UNTIL
SYSTEM COOLS

Freezer will automatically shut
down if critical condenser
temperature is reached to avoid
system damage.

Thermal fuse will automatically
reset after system cools.

HIGH CASCADE
TEMPERATURE

A high cascade condenser
temperature may indicate a
refrigeration problem.

Call New Brunswick Service
department.

HIGH AMBIENT
TEMPERATURE

Room temperature has
exceeded 34 °C. If room
temperature is not reduced, it
may affect freezer performance.

Reduce room temperature

LOW AMBIENT
TEMPERATURE

Room temperature has dropped
below 10 °C. If room
temperature is not increased,
low ambient temperature may
affect performance.

Increase room temperature

8