2 setting high alarm setpoint, 3 setting low alarm setpoint, Setting high alarm setpoint – Eppendorf New Brunswick -86 °C Freezers G User Manual

Page 24: Setting low alarm setpoint, See setting high alarm setpoint on p. 24) , s, Hef® and g -86 °c freezers — operating manual, Op eration

Advertising
background image

HEF® and G -86 °C Freezers — Operating manual

24

6.2.2

Setting high alarm setpoint

The high alarm setpoint may not be warmer than -10 °C and may not be less than within +5° of
the operating temperature. The default setting is +5° from the temperature setpoint.

1. Press the LOCK key.

The LOCK lamp will illuminate, indicating the system is unlocked and parameters can be
changed.

2. Press the HIGH ALARM key.

Its indicator will flash and the display will indicate 0.

3. Using the numerical keys, enter a new alarm setpoint temperature.

The selected temperature will appear in the TEMPERATURE display.

4. When the correct temperature is displayed, press the E (Enter) key to enter the data. To set

the low alarm setpoint skip to (see Setting low alarm setpoint on p. 24), step 2.
The HIGH ALARM indicator will turn off.

5. Press the LOCK key to exit programming.

The LOCK lamp will go off and the freezer will return to normal mode.

6.2.3

Setting low alarm setpoint

The low alarm setpoint may not be colder than -91 °C and may not be more than within -5° of the
operating temperature. The default setting is -5 °C from the temperature setpoint.

1. Press the LOCK key.

The LOCK lamp will illuminate, indicating the system is unlocked and parameters can be
changed.

2. Press the LOW ALARM key.

Its indicator will flash and the display will indicate 0.

3. Using the numerical keys, enter a new alarm setpoint temperature.

The selected temperature will appear in the TEMPERATURE display.

4. When the correct temperature is displayed, press the E (Enter) key to enter the data.

The LOW ALARM indicator will turn off.

5. Press the LOCK key to exit programming.

The LOCK lamp will go off and the freezer will return to normal mode.

Hint!

The LOCK lamp will flash if a lock code (password protection) is required, (see Changing lock
codes on p. 25)
.

Hint!

Press the C key to clear the display during programming.

Hint!

The LOCK lamp will flash if a lock code (password protection) is required, (see Changing lock
codes on p. 25)
.

Hint!

Press the C key to clear the display during programming.

6

Op

eration

Advertising