8 remote alarm (optional), Remote alarm (optional) – Eppendorf Innova 40R Shaker User Manual

Page 13

Advertising
background image

Product description

New Brunswick Innova 40/40R Shaker

English (EN)

13

Tab. 2-1: Alarm descriptions

2.8

Remote alarm (optional)

The Innova 40/40R Shaker can be equipped with a factory-installed remote alarm component (part number
M1320-8029). When it is hooked up to your relay and receiving equipment, this device will send
notification of an alarm condition to the remote location you choose.

Indication

Description

TEMP

• Indicates that the temperature has deviated more than ±1

°C

from

setpoint after achieving control temperature range.

• After door is opened, alarm will be disabled for 5

min while chamber

recovers to setpoint.

RPM

• Indicates that the speed has deviated more than ±5

rpm from setpoint

after achieving operating speed setpoint.

• After door is opened, alarm will be disabled for 5

min while chamber

recovers to setpoint.

POWER

• Indicates that the unit is powering up (both at normal power-up and after

power interruption); will flash until the control knob is moved.

HRS

• Indicates when timed run is completed.

Advertising
This manual is related to the following products: