Excessive operating noise.

• Load out of balance.

Unload all contents and
reload.

• Loose components in platform,

subplatform, and/or drive
assembly.

Call for service.

Shaker does not reach set
temperature.

• Shaker is running in Program

mode (look for Run icon in
display).

• Shaker speed has been

changed by RS-232 command/
computer interface.

• Heater fuse blown.

• Compressor fuse blown.

Replace fuse.

• Compressor over-pressure

switch activated.

• Defective heater.

• Defective refrigeration system.

Call for service.

• Ambient temperature too high

or too low.

Adjust the room temperature.

Incorrect temperature indication.

• Temperature Offset has been

programmed.

Look for Offset icon in display.

• Defective RTD assembly.

• Defective main board.

Call for service.