Product description

New Brunswick Innova 40/40R Shaker

English (EN)

11

Temperature control

Ambient temperature is measured at 1

M from the exterior of the unit.

• Innova

40R provides temperature control from 15

°C below ambient temperature to 80

°C with a

minimum setpoint of 4

°C.

• Innova

40 provides temperature control from 5

°C above ambient temperature to 80

°C.

Both these ranges depend on relative humidity and other ambient factors, as well as the options installed
on the device.

Safety

• Drive Interrupt shuts off power to shaker when lid opens.

• Acceleration/deceleration circuit prevents sudden starts and stops, minimizing both splashing and

mechanical damage.

• Independent mechanical sensing switch also shuts the motor off in unbalanced condition.

Alarms

Innova 40/40R Shaker is equipped with visual and audible alarms that alert the user to the following
conditions:

• The end of a timed run.

• Deviations from speed setpoint.

• Deviations from temperature setpoint.

• Power failure.

• Lid is open.

Platforms

To accommodate customer needs, a wide variety of platforms can be used with the Innova 40/40R Shaker
(see

Platforms on p. 53):

• Universal platforms are the most flexible, providing hole patterns for flask clamps, test tube racks and

other accessories.

• Dedicated platforms are supplied with 1 size flask clamps attached.

Flasks/Tubes

Erlenmeyer flasks (up to 3

L), and a wide variety of tubes and plates can be accommodated using the New

Brunswick shaker accessories (see

Accessories on p. 53).

Other accessories

Test tube racks, microplate holders, and test tube rack holders as well as sticky pads and stick tape are also
available (a universal platform is needed for all test tube racks and holders) (see

Accessories on p. 53).