3 edit a program, 4 run and stop a program, 12 program the timer – Eppendorf Innova 40R Shaker User Manual

Page 35: 13 power interruption, Edit a program, Run and stop a program, Program the timer, Power interruption

Advertising
background image

Operation

New Brunswick Innova 40/40R Shaker

English (EN)

35

The LCD will display

Process Running — Saving Profile

for a few seconds and then return to the main

PROG

screen.

5.11.3

Edit a program

Use the

Edit

mode to open a program you have already created and saved, in order to modify your settings

following the same procedures.

5.11.4

Run and stop a program

Use the

Run

mode to turn a specific program on. Only one program can run at a time. When you change the

mode to

Run

, the screen will show the

Run

icon (see Fig. 5-2 on p. 22).

To stop a program, change the program mode to

Off

.

5.12

Program the timer

By setting an

HRS

setpoint in the

DISP

or

SUMM

screen, the shaker may be set to automatically stop after a

preset time period of 0

hr

1

min

99

hr

59

min.

If the time is set to

0:00

, the shaker will operate continuously until either the lid is opened or the

Start/Stop

switch is pushed.

5.13

Power interruption

In the event of a power failure, the Innova 40/40R Shaker is equipped with an automatic restart function.
The shaker's non-volatile memory retains all stored information.

If the shaker was in operation prior to the power interruption, the shaker will begin to operate at its last
entered setpoints. The alarm

POWER

will flash on the display, indicating that a power interruption has

occurred. Turn the

Control knob in any direction to acknowledge the visual alarm. The flashing will stop.