2 example: requesting measured values, 3 set commands, Example: requesting measured values – Eppendorf Innova 40R Shaker User Manual
Page 39: Set commands
Advertising
Remote programming
New Brunswick Innova 40/40R Shaker
English (EN)
39
6.2.2
Example: Requesting measured values
Type
RV<CR>.
The shaker echos
RV<CR>
The shaker will report:
•
Arg1 (Speed rpm)
•
Arg2 (Temperature °C)
•
Arg3 (Humidity %)
•
Arg4 (CO
2
%)
•
Arg5 (Grow lamp 0 = Off, 1 = On)
•
Arg6 (UV lamp 0 = Off, 1 = On)
These values will be separated by <
Tab>.
This will look like:
250
29
90
0
0
0
6.3
Set Commands
Code
Meaning
Command format
Return data
CS
Command speed
CS<Space><Arg1><CR>
•
Arg1 = Speed setpoint
CT
Command temperature
CT<Space><Arg1><CR>
•
Arg1 = Temperature
setpoint
Advertising
This manual is related to the following products: