5 report request commands, Report request commands – Eppendorf Innova 40R Shaker User Manual

Page 42

Advertising
background image

Remote programming
New Brunswick Innova 40/40R Shaker
English (EN)

42

6.5

Report request commands

PW

Write Program step to
memory

PW<Space><Arg1><Space>...<Arg7><CR>

Arg1 = Program number
(1 – 4)

Arg2 = Step number (1 –
15)

Arg3 = Temperature
setpoint °C

Arg4 = Speed setpoint
rpm

Arg5 = CO

2

setpoint %

Arg6 = Step

(0 – 99)

Arg7 = Step

min (0 – 59)

Code Meaning

Command
format

Return data

RI

Report software version

RI<CR>

Arg1 = Display module software
version

Arg2 = Control module software
version

RP

Report parameter list (ASCII text header)

RP<CR>

Arg1 = Speed rpm

Arg2 = Temperature °C

Arg3 = CO

2

%

RV

Report measured values

RV<CR>

Arg1 = Speed rpm

Arg2 = Temperature °C

Arg3 = CO

2

%

RS

Report setpoint values

RS<CR>

Code

Meaning

Command format

Return data