Page 22: Default alarm contact

Advertising
background image

Operation
TCA-3 Temperature Monitoring System
English (EN)

22

4.6.2

User accounts

The User Accounts tab allows two levels of system access to be created:

• Administrator - Administrator can change any parameters such as alarm levels, details of the alarm

contacts, etc.

• Reader-only - Reader-only can only view the information from the system, and cannot change any

system parameters.

4.6.3

Default alarm contact

The Default Alarm Contact tab contains a list of E-mail addresses, and/or phone numbers. When enabled,
phone numbers can be called in *sequence rather than all at once by selecting Default List when enabling
an alarm for a sensor.

Alarms will be sent once per event, unless sequential alarming mode is selected.

In *sequential alarming mode, the list will be repeated twice if no user takes responsibility by pressing 5.

If the Administrator forgets their password, contact Eppendorf to have the password reset.