Argos Nova Microcinerator User Manual

Page 16

Advertising
background image

16

Tools required:

Cross headed (Philips) screwdriver

7mm AF (9/32” AF) spanner or socket wrench


1. Remove screws (1).
2. Remove front face (2).
3. Remove the suction feet screws (7) and suction feet (6).
4. Remove screw (8).
5. The base plate (5) can now be removed.
6. Disconnect the sterilizing element (3) wires from the on/off switch (12). Note the

connections.

7. Carefully pull the sterilizing element out pushing the wires through from the

underside.

8. Remove the sterilizing element (3) and dispose of appropriately.
9. Remove the new sterilizing element from it’s packing. Check that there is nothing

inside the element core. Caution: the sterilizing element is fragile, please handle with
care.

10. Remove the M4 hexagonal cowl securing screws (14) using the 7mm AF (9/32” AF)

spanner or socket wrench and remove the cowl (13).

11. Carefully feed the wires through the cowl (13) and tilting mechanism (10) and support

shaft (11).

12. Pull the wires through and fit to the on/off switch (12) terminals. The polarity is not

important.

13. Replace the base plate (5) and refit the suction feet (6) and screws (7).
14. Replace the cowl (13) and secure using the M4 hexagonal cowl securing screws (14).
15. Engage the sterilizing element into the rear of the cowl (13) and re-fit the front face

(2).

16. Re-fit the screws (1).
17. Check that the inside of the sterilizing element is clear of debris.
18. Attach the power cord supplied into the rear of the base unit. Plug the opposite end of

the cord in to a suitable power outlet. NOTE: Use of a 2-pronged adapter IS NOT
recommended.

19. Adjust the burner unit to a comfortable position (see page 7). Always allow your

Nova™ Microcinerator to cool for one hour before adjusting the burner head or moving
to a new location.

20. Flip the rocker switch to “on” to activate the unit. Note, the switch should illuminate

red.

21. The Nova™ Microcinerator will reach 816C/1500F degrees within 10 minutes after the

unit is switched on. The Nova™ Microcinerator can be left on for extended periods of
time, though switching the unit “off” when not in use, will increase the sterilizing
element life.


ATTENTION: DO NOT touch any part of the heating guard or end caps when the
Nova™ Microcinerator is in the “on” position or for one hour after being turned
off!