Circuit diagrams, On 7.0) w – Carbolite RHF Series 3 litre - 35 litre User Manual

Page 12

Advertising
background image


RHF


12

MF07 – 3.10

7.0 C

IRCUIT

D

IAGRAMS

In all RHF models the safety switch in the diagram is a positive-break door switch.

EMC Filters (if fitted): dependent on the model there may be one filter, or more than one fitted in
parallel. The circuit diagram examples do not show multiple filter arrangements.

7.1 Single Phase

7.2 2-phase with neutral

As the following 3-phase diagram, except that one phase is not present.

In some models, before mid-2002, direct safety switches were fitted in the element circuit, rather
than the switch in the coil circuit as shown.

7.3 3-phase with neutral

e

l

e

m

e

n

t

(s)

e

l

e

m

e

n

t

(s)

F3

F1

N

F2

L

PE

e

l

e

m

e

n

t

(s)

Instrument Switch

coil

temperature

controller

thermocouple

overtemp.

controller

thermocouple

safety

switch

SSR

heat

on

F3

SSR

SSR

heat

on

F3

heat

on

F3

Filter (if fitted)

if fitted

contactor

if fitted: 2- or 3- circuit

models

e

l

e

m

e

n

t

(s)

e

l

e

m

e

n

t

(s)

F1

N

L2

L3

F2

L1

PE

e

l

e

m

e

n

t

(s)

Instrument Switch

coil

temperature

controller

thermocouple

overtemp.

controller

thermocouple

safety

switch

SSR

heat

on

F3

SSR

SSR

heat

on

F3

heat

on

F3

Filter (if fitted)

if fitted

contactor

note on (2- and) 3-phase: depending on filter(s) fitted, there may be (2

or) 3 separate neutral wires from the elements to the neutral supply.

Advertising