8.2 Power Settings

The power limit settings (parameter

OP.Hi

) for the RHF models on various supply voltages are as

follows. The figures represent the maximum percentage of time that controlled power is supplied

to the elements. Do not attempt to “improve performance” by setting a value higher than the
correct one from the table.

Model

Volts:

200V 208V

220V

230V

240V

380V

400V

415V

RHF 14/3

90

83

74

68

63

74

68

63

RHF 15/3

62

57

51

47

43

51

47

43

RHF 16/3

62

57

51

47

43

51

47

43

RHF 14/8

90

83

74

68

63

74

68

63

RHF 15/8

60*

56*

76

69

63

76

69

63

RHF 16/8

73*

68*

74

68

62

74

68

62

RHF 14/15

92

85

76

69

64

76

69

64

RHF 15/15

71

66

59

54

49

59

54

49

RHF 16/15

73

68

61

55

51

61

55

51

RHF 14/35

93

86

77

71

65

77

71

65

RHF 15/35

93

86

77

71

65

77

71

65

RHF 16/35 (star)

-

-

-

-

-

85

78

71

RHF 16/35 (delta) -

89

79

73

67

-

-

-

* these models use different elements in 200-208V versions