Fuses & power settings, On 8.1 – Carbolite RHF Series 3 litre - 35 litre User Manual

Page 14

Advertising
background image


RHF


14

MF07 – 3.10

8.0 F

USES

&

P

OWER

S

ETTINGS

8.1 Fuses

F1-F3: Refer to the circuit diagrams.

F1

Internal
supply fuses

Fitted if supply cable fitted.
Fitted on-board to some types of EMC filter.

on-board and up to 16 Amps:
32mm x 6mm type F
other: GEC Safeclip

F2

Auxiliary
circuit fuses

Fitted on-board to some types of EMC filter.
May be omitted up to 10 Amp/phase supply rating.

2 Amps glass type F
On board: 20mm x 5mm
Other: 32mm x 6mm

F3

Heat Light
fuses

May be omitted up to 25 Amp/phase supply rating. 2 Amps glass type F

32mm x 6mm

Customer
fuses

Required if no supply cable fitted.
Recommended if cable fitted.

See rating label for
amperage; see table below
for fuse rating.

Model

phases

Volts

Supply Fuse Rating

RHF 14/3

1-phase

200-240

32A (or 30A)

RHF 14/3

2-phase + N

380/220 - 415/240

16A/ph (or 15A)

RHF 15/3,16/3

1-phase

200-240

40A

RHF 15/3,16/3

2-phase + N

380/220 - 415/240

20A/ph

RHF 14/8

1-phase

200-240

63A

RHF 14/8

2-phase + N

380/220 - 415/240

32A/ph (or 30A)

RHF 15/8,16/8

1-phase

200-240

63A

RHF 15/8,16/8

3-phase + N

380/220 - 415/240

20A/ph

RHF 15/8,16/8

3-phase delta

220-240

32A/ph

RHF 15/8,16/8

3-phase delta

208

40A/ph

RHF 14/15

3-phase + N

380/220 - 415/240

25A

RHF 14/15

3-phase delta

220-240, 208

40A/ph

RHF 15/15,16/15

3-phase + N

380/220 - 415/240

25A/ph

RHF 15/15,16/15

3-phase delta

220-240, 208

50A/ph

RHF 15/15,16/15

3-phase + N

380/220 - 415/240

25A/ph

RHF 15/15,16/15

3-phase delta

220-240, 208

50A/ph

RHF 14/35, 15/35

3-phase + N

380/220 - 415/240

40A/ph

RHF 14/35, 15/35

3-phase delta

220-240, 208

63A/ph

RHF 16/35

3-phase + N

380/220 – 415/240

40A/ph (see note)

RHF 16/35

3-phase delta

220-240, 208

63A/ph (see note)

380-415V 3-wire: same ratings as 3-phase+N

RHF 16/35 – technical note.

All phases should be fused as indicated: the same wire size is used in all three internal circuits. For the
delta configuration the furnace power is set lower to bring the rated current below 63 Amps. The current
ratings for the three phases are:

380/220 – 415/240

208-240 delta

Furnace Power

16000W

15000W

Phase 1

40 Amps

63 Amps

Phase 2

40 Amps

48 Amps

Phase 3

20 Amps

48 Amps

Advertising