Fault analysis – Carbolite CDF 15/1b User Manual

Page 12

Advertising
background image


CDF


12

MF56 – 1.03


New furnace: the diagrams indicate the connections. If you have any problems with this procedure,
please contact our service division. A common feature of CDF models is that elements are in
series between 200-240 volts.

Replacements: make a note of the actual braid and cable connections to the elements. Using the
reverse method described above, disconnect the braids or cables and carefully withdraw each
element.

Carefully insert the new elements into the furnace and attach the braids according to the notes
made, or according to the appropriate diagram below

Whether fitting new elements or replacing old, replace the furnace back panel when connections
are complete remembering to connect the earth cable to the back.

Element connections – 3 Elements

Photograph shows typical live and neutral connections

L1

N