2 save snapshot, 3 load configuration, 2 save snapshot -4 12.3 load configuration -4 – CiDRA SONARtrac GVF-100 User Manual

Page 106

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 12-4

20675-01 Rev 02

12.2 Save

Snapshot

Save Snapshot

is a one step download of a pre-determined data set

from the transmitter. The data set is comprised of the following.
(Download of Snapshot takes about 6 – 7 minutes.)

Raw Data

Raw Data is as the name implies, a set of unprocessed data from the
sensor head.

A raw data file (5 minutes duration) is downloaded during Snapshot.

Data History

Data History

is a compilation of flow system information that is saved

within the transmitter. The data history file is a text file that contains
the data output from the transmitter as well as some system
information that can be used for diagnostic purposes. The duration
(number of days) of data stored in data history is primarily determined
by the system data update rate. Under default setup (2 second
update rate) approximately 24 days of data will be stored.

A data history file of the last 24 hours is downloaded during Snapshot.

System Info

System Info

(information) is a list of system hardware and software

revisions and versions, system diagnostic information, and system
setup information.

Configuration

Configuration

is a full list of the transmitter setup parameters. The

file is encrypted for security purposes.

Event Log

Event Log

lists a history of the abnormal operations within the system

that may affect its performance. Up to 23 events are stored in this file;
after that the oldest event will be overwritten with a new event.

12.3 Load

Configuration

Load Configuration

is one-step command used to upload a

configuration file to the transmitter from the SONARstick

TM

.

In the Load Configuration mode the transmitter will only load files with
the serial number of the transmitter followed by ’_LoadConfig.ini’ in the
file name. For example, a configuration file for transmitter serial
number 234 must be named ‘234_LoadConfig.ini’ for it to be uploaded
by this utility.

Advertising
This manual is related to the following products: