CiDRA SONARtrac GVF-100 User Manual

Page 69

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 9-18

20675-01 Rev 02


Table 14 Transmitter Menu Tree Software Release 03.03.XX (page 3)

Level 1

Level 2

Level 3

Range

Description

Output Sel

Flow Rate, SOS, GVF,
Blank

Parameter to output

Power Sel

Internal, External

Power supply for 4-
20mA current

Low End

Based on ‘output sel’

Low end output (4mA)

High End

Based on ‘output sel’

High end output (20mA)

Out of Range

Hold, >20mA, <4mA,
4mA

Behavior when meter
reading is out of range
or meter has no reading

Overrange
Rail

Enable, Disable

‘Enable’ causes the
output to go to full
(20mA) or min (4mA)
when system over
ranges

4mA Trim

2 to 6

Adjust 4mA output

4-20mA
Ch1& Ch2

20mA Trim

18 to 22

Adjust 20mA output

Multiplier

0 - 999999

Multiplier on output

Width (ms)

.5,1,20,33,50,100

Pulse width

Lowcut

0 % to 100 %

Low end cutoff

Pulse

Output Sel

Flow Rate, Totalizer,
Flow Rate%, GVF, SOS

Parameter to output

Warning Off/On

Activates alarm warning
function

Critical

0 % to 100 % Off/On

Activates alarm critical
function

Alarm
Control

Manual Clear Off/On

Allows for manual or
automatic alarm
clearing

Flw Min

Flw Max

GVF Min

Alarm
Warning
Threshold

GVF Max

0-100% of range

Used to set the
minimum and maximum
values for a warning
alarm

Flw Min

Flw Max

GVF Min

Output Config

Alarm Critical
Threshold

GVF Max

0-100% of range

Used to set the
minimum and maximum
values for a critical
alarm

Table 14 Transmitter Menu Tree Software Release 03.03.XX (page 4)