9.5 Transmitter

Display

The transmitter display functions in two distinct modes: an operational
mode and a menu mode. These two modes will be explained in the
following sections.

9.5.1 Operational

Mode

The transmitter screen displays the status of the SONARtrac

TM

system. A few typical screen messages and their interpretation follow.

9.5.1.1 System

Start

The transmitter displays the startup progress as it loads firmware into
various parts of the system and the Ethernet IP address of the
transmitter as it starts up. Once startup is complete, the display will
clear and begin displaying measurements. Full system startup is
discussed in Section 10.

Figure 51

VF & GVF Startup Screen

If internal configuration parameters are changed in the menu screens
once the menu is exited, the display will clear and show the ‘Startup
>>>>’ display. This indicates that the system is restarting with the
new parameters and is beginning to make measurements. Flow and
GVF measurements will be displayed when the startup procedure is
complete.

Startup >>>>>>>>>>>>>

010.001.009.13

D 01.02.10 P 01.00.14 /

System is
starting

Firmware
revisions as
they load:
I: Interface
F: FPGA
D: DSP
P: PreAmp

Ethernet IP
Address

Mode and
heartbeat

Advertising