power supply

230/240 V 50/60 Hz / 115 V 50/60 Hz (optional)

temperature range

5 K above ambient (65°C max.)

heating power

300 W

accuracy (temperature
control)

± 2 K up to 50°C
± 4 K above 50°C (air temperature)

temperature distribution

± 2 K up to 50°C
± 3 K above 50°C

time to heat circulating air

abt. 12 minutes from ambient to 50°C
abt. 25 minutes from ambient to 60°C (unobstructed
circulation)

protection class

IP 30

duty cycle

100 %

temperature control

electronic

temperature sensor

Pt 100, installed in heater module

heater

tubular heater with overheat protection

circulation of air

cross current fan, recirculating principle

temperature display

nominal: 2-digit LED display
actual: 2-digit LED display

size (w x d x h)

abt. 170 mm x 375 mm x 97 mm



Hood for Incubator 1000

material

PETG, split-design

size (w x d x h)

abt. 570 mm x 390 mm x 270 mm

volume (air)

abt. 50 dm³


Hood (flat) for Incubator 1000

material

PETG, split-design

size (w x d x h)

abt. 570 mm x 390 mm x 170 mm

volume (air)

abt. 30 dm³