Kds model 200 series, Operating instructions – KD Scientific 200 Series User Manual

Page 13

Advertising
background image

9

KDS Model 200 Series

Operating Instructions

(Continued)

SYRINGE DIAMETER ENTRY

The pump must be calibrated by identifying the internal diameter of the syringe used.

Once entered this data is stored in non-volatile memory and need be modified only when a

different syringe is used.

Table

Use the arrow keys to move TABLE to the center of the display and press select

to select this feature. The display now reads abbreviated names of syringe

manufacturers and the type of syringe (plastic or glass).

Use the arrow and select keys to scroll through the list and select the manufacturers’

name of the syringe used.

The display now lists the size of syringe. Again, use the arrow and select keys to

identify and enter the syringe used.

Diameter

If the syringe used is not listed in the table of syringes then the internal diameter of the

syringe must be measured and entered directly.

Scroll through the main menu and select DIA.

The display now prompts for entry of the syringe barrel internal diameter measured in

millimeters. Enter by using the numerical keys and enter.

Note: If the diameter is changed the volume and rate settings are set to zero.

VOLUME SETTING

Volume can be reached directly from the main menu or will be prompted after mode selection.

Volume prompt: Vol.: 00.00ml >

1)

Enter target volume from the numerical keypad.

2)

If the units and value displayed are correct, enter.

The underlined display, or pulsing display segment, indicates that this parameter can be

changed. The arrow symbol signifies direction of travel (< indicates withdrawal ) and pulses

when the pump is running.

Note: When dispensing, the displayed volume increments in units of the last significant

figure of the volume entered. Therefore, to increment by .01 the volume should be set at

1.00. If the volume is set at 1.000 then the volume increments in units .001.

Note: Volume setting = 0.0

a)

Infusion or Withdrawal modes.

Volume = 0 is interpreted as no volume and the pump will run until manually

stopped or a stall occurs. The LCD will display the rate setting.

b)

Infusion/withdrawal, Withdrawal/infusion and Continuous modes.

A volume setting must be entered. If no volume is entered the display returns to

the mode selection menu to select the appropriate mode.