KDS Model 200 Series

Specifications

Model

200 Series

Syringe size

10 microliter to 140 milliliter

Electrical Rating

US model 115 V~, 0.25 A
CE model 230 V~, 0.16 A

Fuse

5 x 20 mm, 250 V~, Slow Blow, 0.25 A

Voltage operating range

US model 110-120 V~, 50/60 Hz
CE model 200-240 V~, 50/60 Hz

Drive mechanism

Microprocessor controlled stepper motor

1/2 - 1/16 microstepping, driving a leadscrew

through a belt and pulley drive mechanism

Pusher advance per microstep

(1/16 step) - 0.165 micron (or .0000064 inch)

Volume per microstep

(1/16) with 60ml BD syringe - 0.0919µl

Max stepping rate

1600 (1/2 step) /sec

Min step rate

1 (1/16 step) /120 secs

Linear Travel Rate

Min: 4.95 x10

-4

cm/hr

Max: 12.67 cm/min

Flowrate range

2.757 µl/hr to 70.56 ml/min (60 ml syringe)

Nominal linear force

35 lb

Dimensions

11x9x5.5 inch (28x23.5x14 cm)

Weight

9 lb (4 kilo)

Atmospheric Specifications

Temperature

5˚C - 40˚C (41˚F - 104˚F)

Humidity

20% - 80% RH

Mode of Operation

Continuous

Classification

Class I

Pollution Degree

2

Installation Category

II

Output

N/A

Physiological Effects

N/A

Cooling Conditions

No special considerations

Mechanical Stability

No special considerations

Protective Packaging

No special considerations

Earth Terminals

No External connections required

Removable Protective Means

N/A

Supplier Name

KD Scientific Inc.

Address

84 October Hill Road, Holliston, MA 01746-1388