KD Scientific Legato OEM 950 User Manual

Page 13

Advertising
background image

13

Operating the KDS Legato 950 Series

K

D

S

L

e

g

a

to

9

5

0

O
E

M
S

e

ri

e

s

U

s

e

r’s

M
a

n

u

a

l

KD Scientific Legato 950 OEM Series Syringe Pump User’s Manual

Publication 5617-012 Rev A

Argument errors

Argument errors are displayed when a command argument

is unrecognized or out of range. The argument in question

will be displayed except in the case of missing arguments.

The argument error has the following format:

<lf>[##:]Argument error: [bad argument]<cr>

<lf>[##:]<sp><sp><sp>{error message}<cr>

<lf>[##]<prompt>

System commands

ADDRESS

Sets or displays the pump address. Valid range is 0 to 99.

Command format:

address [0-99]

Query response:

<lf>[##:]Pump address is #<cr>

<lf>[##]<prompt>

CATALOG

Displays a catalog of programs stored in the pump.

Command format:

cat

Query response:

<lf>[##:]Program name Size<cr>

<lf>[##:]———————- ——<cr>

<lf>[##:]{program name} #<cr>

This line is repeated for each file on the disk

<lf><lf>[##:]# file(s) using # steps<cr>

<lf>[##]<prompt>

The last line shows the number of files and the total number

of steps being used on the disk.

NOTE: Pumps with an address of 0 are

masters, and pumps with an address

between 1 and 99 are slaves.

NOTE: The file size is the number of steps in

the program plus one for the file information

header.

CONFIG

Sets or displays pump configuration.

Command format:

Config [{config designation}|{value}]

DELMETHOD

Deletes a program, stored in the pump.

Command format:

delmethod {program}

ECHO

Sets or displays the USB echo state. Valid states are on

or off.

Command format:

echo [on|off]

Query response:

<lf>[##:] OFF<cr>

<lf>[##]<prompt>

or:

<lf>[##:] ON<cr>

<lf>[##]<prompt>

FORCE

Sets or displays the infusion force level in percent. Valid

range is 1 to 100.

Command format:

force [1-100]

Query response:

<lf>[##:]#%<cr>

<lf>[##]<prompt>

FTSWITCH

Sets or displays the footswitch setting. The footswitch

may be used in a toggle on or off mode (momentary), a

press to run mode (falling), or a release to run mode

(rising).

Command format:

ftswitch [mom|rise|fall]

Query response:

<lf>[##:]Momentary<cr>

<lf>[##]<prompt>

or:

<lf>[##:]Active high<cr>

<lf>[##]<prompt>

or:

<lf>[##:]Active low<cr>

<lf>[##]<prompt>

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

NOTE: After changing the configuration, the

pump must be power cycled.

Advertising