TROUBLESHOOTING

1. Infusion Accuracy

To ensure infusion accuracy always use new syringe(s)

and measure syringe bore diameter and enter actual

dimensions in millimeters (mm) using the Custom Syringe

entry option. Additionally, make sure that the guide rods

and lead screw are properly lubricated.

KDS Legato 950 OEM Series Display Error

Messages

2. Motor Stalled

This indicates that the pusher block travel has been

impeded. This may be caused by the syringe plunger

hitting bottom, a kink in the tubing (occlusion), syringe

plunger binding or any situation requiring more force to

the head of the syringe than the pump is capable of

delivering. To restart the pump, remove the cause of

stalling and/or increase the force setting, then press

RUN. The pump will continue the interrupted procedure

where it stopped.

3. Out of Range

A value was entered or encountered in a pump program

that was beyond the pump’s limits.

4. Unit Won’t Power Up

Verify the the Power Source Switch on the rear of the

pump is set to DC-IN if powering through power supply

or RS-485 if powering in a chain.