Leica CM1860/CM1860UV – Cryostat

6.2.2 Setting the automatic defrost time (cryochamber)

The automatic defrost cycle takes place once every 24 hours.

Briefly touch the or button to display the beginning of the defrost cycle

which has currently been set. The LED's between the indication of hours

and minutes will flash.

To change the beginning of the defrost cycle in 15-minute increments, press

the or button. The defrost time is 12 minutes.

6.2.3 Selecting the cryochamber temperature

The temperature of the cryochamber is set and indicated on the panel marked

with the cryostat symbol.

The actual temperature is the standard indication. Briefly touching the

or the button displays the target temperature.

You can now set the desired value using these buttons. Pressing the or

button for more than 1 sec. increases or decreases the chamber tem-

perature continuously.

The actual value will be indicated 5 seconds after finishing the programming.