Daily use of the instrument – Leica Biosystems CM1860/CM1860 UV User Manual

Page 47

Advertising
background image

47

Leica CM1860/CM1860UV – Cryostat

Knife holder CN with anti-roll guide – Moving the clamping chucks

2

3a

4

4

3

The clamping jaws are factory-installed in the

knife holder with a clearance of 64mm. If neces-

sary, both clamping jaws can be offset with a

clearance of 84 mm.
Perform the following steps:
• Use a 4 mm Allen key to loosen the screw

over the clearance angle adjustment (23) and

remove the segment arc (2) from the blade/

knife holder base.

• Use a 4 mm Allen key to loosen the screws (4)

on the underside of the segment arc.

Fig. 46

5

6

Fig. 47

• Lift off the clamping jaw (5) on the right side

(caution: washers) and feed it into the adjacent

hole (6). Tighten the screws from the underside

of the segment arc. Repeat on the left side.

Now use the longer knife back rest provided.

If large specimen discs are to be used in the specimen holder (e.g. 50 x 80 mm), it is pos-

sible to shift the clamping jaws laterally.

3a

Never work with only one clamping

jaw, as this does not ensure the stability

required for the sectioning process.

Also, a long knife will not be sufficiently

protected by the knife guard in this

instance.

23

7. Daily Use of the Instrument