31

Leica RM2125 RTS

Prior to sectioning, laboratory person-

nel MUST check that the cassette is

seated securely in the universal cas-

sette clamp.

• Push the lever (60) forwards.
• Mount the cassette (61) horizontally or verti-

cally as required.

• To clamp the cassette, release the lever.

5.9.5 Universal cassette clamp

(optional)

Fig. 24

60

61

5. Operation

Leica / Surgipath cassettes with minimum dimensions of 39.8 x 28 mm and maximum dimen-

sions of 40.9 x 28.8 mm may be clamped in the universal cassette clamp (UCC) horizontally as

well as vertically.
When using other – particularly thin-walled – cassettes, the cassette may become deformed

or other problems may arise with the clamping system. If the user tries clamping the cassette

and realizes that it is not securely clamped into place, an alternative tensioning clamp must be

used.
When using cassettes whose lid is molded on, make sure that the broken edge left by removing

the lid does not prevent the specimen from being securely clamped – if necessary, the speci-

men must be clamped horizontally.
Before clamping the cassette into the universal cassette clamp, remove excess wax on the

outside of the cassette to ensure that the cassette clamps in securely.
Wax deposits on the outside of the cassette can make the universal cassette clamp dirty. The

dirt prevents the cassette from clamping in securely and can lead to sections being too thick

or thin, chattering within the section and, in the worst-case scenario, damage to the specimen.

Prior to sectioning, the user has to verify that the specimen is clamped securely and, if neces-

sary, remove wax deposits from the universal cassette clamp according to the specifications in

Chapter 8.1 "Cleaning and maintenance – Universal cassette clamp".