3 integrated safety devices, Integrated safety devices, Safety – Leica Biosystems RM2125 RTS User Manual

Page 9

Advertising
background image

9

Leica RM2125 RTS

Warnings – Maintenance and Cleaning

• Only authorized and qualified service personnel may access the internal components of the

instrument for service and repair!

• Do NOT catch accessories that are falling down – there is a risk of injury!

• Lock the handwheel before cleaning!

• Do not use any solvents containing acetone or xylene for cleaning!

• Ensure that no liquids enter the interior of the instrument when cleaning!

• When using detergents please comply with the safety precautions of the manufacturer.

2.3 Integrated safety devices

The instrument is equipped with the following safety devices:

2. Safety

Handwheel lock
The handwheel (15) can be locked in the

12 o'clock position (

Fig. 2

).

As soon as the handwheel handle (1) is pushed

in to the left, the handwheel engages the next

time it reaches the 12 o'clock position, which

locks it mechanically.
Functional test:

• To activate the lock, press the handwheel

handle (1) to the left. The handwheel is now

mechanically locked in the 12 o'clock posi-

tion and can no longer be rotated.

• To deactivate the lock, pull out the hand-

wheel handle (1) to the right.

Handwheel brake
Using the lever (2) on the right side of the mi-

crotome base plate, the handwheel brake can

be activated with the handwheel in any posi-

tion. If the lever is pressed upwards, the hand-

wheel cannot be moved any more. The two le-

ver positions are marked by corresponding dots

on the microtome base plate (

Fig. 2

).

Fig. 2

Lever in position  = handwheel locked

Lever in position

= handwheel released

1

2