Cleaning and maintenance – Leica Biosystems RM2125 RTS User Manual

Page 43

Advertising
background image

43

Leica RM2125 RTS

Universal cassette clamp

• Detach cassette clamp (13) for a thorough cleaning, removing all

paraffin residues.

• For cleaning, do not use xylene. Use xylene substitutes or paraffin

removers such as "Para Gard."

• The cassette clamp (13) can also be placed in a drying chamber heated

to a maximum of 65°C, until the liquid wax escapes.

There is a burn hazard when removing the parts from the drying

chamber (65 °C). Wearing safety gloves is recommended!

• Remove paraffin residues with a dry cloth.
• After such a cleaning procedure in an oven, always be sure to lubricate

the axle and spring of the clamping lever (60) (see also

Chapter 6.2

).

6. Cleaning and Maintenance

Fig. 36

60

13

• Reassembly takes place in reverse order.
• When performing the installation, make sure that the upper edge of the

pressure plate (86) is parallel and level with the upper edge of the rear

pressure plate (80) (

see also Fig. 30/31

, page 35). If necessary, adjust

the pressure plates (

Chap. 5.9.10

).

90