Control test, Prepare for the test – Bayer HealthCare Elite XL User Manual

Page 12

Advertising
background image

8

Glucometer Elite

®

XL

section

Prepare for the Test:

1. To open the Test Strip foil packet,

carefully peel the foil back until
the Test Strip is completely
exposed (as shown). Save the
foil packet.

Note: Numbers printed on each Test Strip
are manufacturing numbers and are not to
be confused with Code Numbers (F#).

2. Holding the round end of the

Test Strip, insert the strip into the
Meter until it stops.

3. A beep sounds. A full display, fol-

lowed by the Code Number (F#)
and previous test result, appears.
The Code Number and previous
test result flash alternately.

4. To mark a Control Test, pull the

Test Strip out and reinsert quickly
(within 2 seconds). When the “C-”
and Code Number (F#) begin
flashing, you know the test will
be marked.

Control Test

A Control Test is run to find out if the
whole
system is working correctly (Test
Strip and Meter) and
to practice good
technique before blood glucose testing.

5