Test Sample: Whole blood

Test Result: Plasma/serum glucose

Reaction Chamber

Sample Volume: 2

␮L

Measuring Range: 20–600 mg/dL

(1.1–33.3 mmol/L)

Measuring Time: 30 Seconds

Date: Year 2000 compliant

Calibration Curve: Selected by use of the

Code Strip

Memory Feature: Stores last 120 test results

Temperature

Compensation: Automatic compensation

with built-in thermistor

Battery Type: 3-volt lithium batteries

(DL or CR2032) – 2

Battery Life: Approximately 1000 Tests

(1 year of average use)

Operating

:

Temperature Range: 50°–104° F (10°–40° C)

Humidity: 20–80% RH

Dimensions: 97.8 x 56 x 14.5 mm

(3

7

8

x 2

3

16

x

9

16

in.)

Weight: Approximately 60g

(2.1 ounces)

Alarm: “Beeps” whenever a Check

Strip, Code Strip or Test
Strip is inserted into the
Meter, when the Test Strip
is filled with blood or
Control Solution, and
when a test result appears
in the display.