MESSAGE

CAUSE

SOLUTION

2I

Glucometer Elite

®

XL

II

section

Meter fails to start
countdown.

1. Blood applied

wrong; see page
12.

2. Test Strip

inserted
incorrectly.

3. Meter turned on

before Test Strip
inserted.

4. Sample is applied

while full display
is showing.

Some display seg-
ments do not appear.

Meter electronics or
battery failure.

Replace batteries and if display
is still incorrect, contact
Customer Service Department.

Battery display.

Battery voltage
low.

Replace batteries. See
Section 1, page 4.

1. Form a larger drop of blood

or Control Solution and
rerun test using a new Test
Strip. Hold test end to sam-
ple until after the “beep”.

2. Insert Test Strip correctly

(Meter end first).

3. Press and hold button to

turn Meter off. Remove
Test Strip and reinsert to
begin test procedure.

4. Wait for Code Number

(F#) and last test result
to begin flashing before
applying sample to strip.