CELSIUS 10 FDA-Edition

8

9

Time

hh:mm

Period

hh:mm

Flap

Type

Fan

Edit Profile

2

1

3

4

5

6

7

Tabular View (example)

1 Ramp segment No. ...

2 Beginning of ramp segment after programme start

3 Duration of the ramp segment

4 Temperature in the ramp segment

5 Ramp close statement (see page 30 )

30

30

6 Flap position

7 Fan speed

8 Position of switching contact a

9 Position of switching contact b

In addition to the ramp segments entered, the table always shows an empty line (in the exam-

ple: line 8). A further ramp segment can be added here.

Inserting and deleting lines

Inserting and deleting lines

If the table window is active, the commands for deleting and inserting lines are additionally

available in the view menu and toolbar:

Insert a line at the position marked

Delete selected line

If the last ramp segment is deleted, the ramp close statement is automatically transferred

to the line above.