Cyclic Command Process

MTS MultiPurpose TestWare®

Command Processes

191

Cyclic Command Parameter Descriptions

Command tab

Command Tab (part 1 of 2)

I

TEM

D

ESCRIPTION

Segment Shape

Specifies the segment shape.

Time, Frequency, Rate

Specifies how fast each segment runs.

The following methods are supported:

Time specifies the time to run one segment.

Frequency specifies the time to run a two-segment cycle.

Rate specifies a constant rate between the starting level and the end
levels.

Note

You cannot use Rate for a command process applied to multiple
channels. If you do, an error message will appear when the procedure
is locked.

Note

The rate of the first segment can be faster than the other segments if it
must transition from a distant end level.

Count

Specifies how many times the waveform will be repeated.

Note

If you clear the Count box, the waveform will repeat continuously.

Adaptive
Compensators

Enables command compensation during the command process.

Important

The use of the AICX compensator can introduce large delays
between generated command and command to the actuator.
The delay is dependant on the tuning of AICX and is two times
the value of the Impulse Response.

Do Not Update
Counters

Disables the counters. When enabled, the Channel Counters do not increment

for this process and the counter totals are unaffected.

Relative End Levels

Defines the segment end level:

Absolute End Level is relative to setpoint.

Relative End Level is relative to the starting command value (the value
when the process begins).

Channel

Specifies the channel.

Control Mode

Specifies the control mode or control-mode variable.