MTS Series 793 Application User Manual

Page 174

Advertising
background image

MTS MultiPurpose TestWare®

Set Process Parameters in the Group Table

Creating a Practice Test Procedure

174

4. Define the second Data Limit Detector process (Trigger at 50C).

A. In the group table, click the Data Limit Detector (Trigger at 50C)

process icon.

B. Click the Signals tab and move the Temperature Monitor resource to

the Included list, as shown below.

C. Click the Limits tab. Type and select the parameters needed for the

test, as follows:

Signals: Temperature Monitor.

Upper Limit (enabled): 50 degrees C.

Lower Limit (disabled)

D. Observe the default settings on the Settings, Limit Tables, and

General tab.

Data Limit Detector
(Trigger at 50C)