Peak/Valley Acquisition Process

MTS MultiPurpose TestWare®

Data Acquisition Process Descriptions

279

Peak/Valley Acquisition Process

About Peak/Valley Process

The Peak/Valley Acquisition process records the output of all selected signals
when the software detects a peak or valley in the master signal you specify.

When you define a Peak/Valley Acquisition process, you must define the master
signal, the sensitivity, the signals you will record data on, the data destination, the
buffer type, and the output units.

Sensitivity

A sensitivity value specifies how much the master signal must change to detect a
peak or valley:

Setting the sensitivity too low may cause signal noise to be recognized as
peaks and valleys.

Setting the sensitivity too high may cause low amplitude signals to be
missed.

Acquired Data Points

Sensor
Signal

Time

Detected
Peaks

Detected
Valleys

Sensitivity =