Peak/valley change detector process example – MTS Series 793 Application User Manual

Page 367

Advertising
background image

Peak/Valley Change Detector Process

MTS MultiPurpose TestWare®

Event Process Descriptions

367

Peak/Valley Change Detector Process Example

Suppose you want to acquire a cycle of data each time the peaks or valleys of an
input signal change by a specified amount. Further, suppose you define a Peak/
Valley Change Detector
process with the following parameters:

Trigger continuously

Tolerance: ±50 lbs

Sensitivity: 5 lbs

The test command cycles between ±0.0001 strain and you are monitoring a force
signal.

Each time the peaks or valleys on the force signal change more than ±50 lbs, the
process generates a trigger signal that starts a data acquisition process.

The data acquisition process would use the Peak/Valley Change Detector
process as its start trigger. A single data acquisition process could be run multiple
times with this change detector.

Note

It is good practice to specify a number of cycles to run before you begin
monitoring to allow a “settling in” period.

Strain Command

100 lbs (

±

50 lbs) Bands

Trigger Data Acquisition Processes

Force Signal

Advertising
This manual is related to the following products: