Program control parameter descriptions – MTS Series 793 Application User Manual

Page 401

Advertising
background image

Program Control Process

MTS MultiPurpose TestWare®

Other Process Descriptions

401

Program Control Parameter Descriptions

Action tab

Action Tab (part 1 of 3)

I

TEM

D

ESCRIPTION

Action

Specifies the action that occurs when the process completes.

None

The process ends without generating a message or performing an action.

Message Only

The process writes a message to the MPT Specimen Log. (You define the

content of the message in the Message box.)

Program Hold

The process suspends the test until you click the Run button on the Station

Controls panel. This action functions the same as clicking the Hold button on

the Station Controls panel.

Program Stop

The process stops the test. This action functions the same as clicking the Stop
button on the Station Controls panel.

Program
Pause

Similar to the Program Hold action, but the test must be resumed by a

Program Resume action from another Program Control process.

Program
Resume

Resumes a test paused by a Program Pause action.

Procedure
Error

The process stops the test and generates an error. The procedure goes to the

error state and cannot be resumed. To run the procedure again, click the Reset

button, then click the Run button on the Station Controls panel.

Station Power
Off

The process removes hydraulic power from the station. The test can be

resumed after you reset interlocks and reapply hydraulic power to the station.

Interlock

The process triggers a hydraulic interlock and shuts down the hydraulic power
supply. The test can be resumed after you reset the interlock, reactivate

hydraulic power, and apply hydraulic power to the station.

Program
Interlock

The process triggers a program interlock. The test can be resumed after you
reset the interlock.

Advertising