Model 532 Configuration

Press the EZ-Dialä Knob while turning the Model 532 on to access the

configuration mode. Turn the EZ-Dialä Knob to select configuration items.

Press the EZ-Dialä Knob to change configuration items. Turn the unit off or

just wait approximately 8 seconds to exit the configuration mode.

Auto Off -

ON (default)/OFF

Auto Off is ON, by default, to save battery life by turning the unit off after 30 minutes of inactivity. Turn

Auto Off to OFF to prevent automatic shutdown. This is typically useful for manual loading or continuous

use.

EZ-Step

ä -

ON/OFF (default)

If EZ-Stepä is ON manual and automatic stepping/ramping is available. If EZ-Stepä is OFF the EZ-Stepä

pushbutton will be disabled and the step direction indicator will not be displayed.

HART® Compatibility Mode -

ON/OFF (default)

The Model 532 has a HART® compatibility mode. This mode is useful when the devices being powered

communicate using the HART® protocol. In this mode the Model 532 connects a 250 Ω load resistor in

series with the output in both Source and Power Measure 2-Wire transmitter modes. This eliminates the

requirement of an external 250 Ω load resistor. This resistor is typically shown in connection diagrams and

manuals for HART® devices.

If HART® Compatibility Mode is ON, a 250 Ω load resistor is automatically switched in series with the

output in Source and Power Measure 2-Wire Transmitter modes. The output compliance with HART®

Compatibility Mode ON is 950 Ω at 20 mA.

If HART® Compatibility Mode is OFF there is no 250 Ω load resistor in series with the output. This will

increase the output compliance voltage to drive 1200 Ω at 20 mA.

EZ-Check

ä HI/LO Readings

ON/OFF (default)

If the EZ-Checkä HI/LO Readings option is ON, the highest and lowest readings will automatically be

saved in the HI and LO EZ-Check™ positions.

If this option is OFF the HI and LO positions will show the current reading.

Loop Diagnostic

ON (default)/OFF

Loop Diagnostic may be turned off to prevent entry into diagnostic mode. This may make the Model 532

simpler to operate by eliminating accidental entry into diagnostic mode.

Factory Reset

ON/OFF (default)

If Factory Reset is ON, the unit will restore all factory defaults when the Model 532 is turned OFF and back

ON. This will reset any changes made in the Model 532 Configuration options, returning the unit to its

simplest factory configuration.