Source Milliamp

Connect the Model 532 directly to 4-20 mA

receiver equipment, alarms, panel meters,

etc. Use the EZ-Dialä Knob and EZ-Checkä

Switch to adjust loop current. The display

flashes “HIGH Ω” when the loop resistance is

too high or the leads are open.

2 Wire Transmitter Simulate

Substitute the Model532 for a 2 wire

transmitter. Use the EZ-Dialä Knob and EZ-

Checkä Switch to adjust loop current. At

least 2 volts of loop power is required, else

the display flashes “CHECK LOOP SUPPLY.”

Source Volts

The Model 532 sources 0.000-24.000 volts.

This is useful for powering transmitters and

receiver equipment. Use the EZ-Dialä Knob

and EZ-Checkä Switch to adjust output

voltage. The display flashes “LOW W” when

the output is overloaded.

K.

Warranty

Our equipment is guaranteed against defective material and workmanship (excluding batteries) for a period of three years from the
date of shipment. Claims under guarantee can be made by returning the equipment prepaid to our factory. The equipment will be
repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our
guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment. Under no
condition shall Practical Instrument Electronics, Inc. be liable for any special, incidental or consequential damage.

Advertising