Auto Step/Ramp

Press the EZ-Stepä pushbutton for more than one second to activate auto step/ramp mode. The Model 535 will automatically step by

the selected EZ-Stepä size and time. Press the EZ-Stepä pushbutton again to deactivate auto step/ramp mode.

The EZ-Stepä direction is indicated on the display (

or

). Press the EZ-Dialä knob to change the step direction. The step

direction can be changed while automatically stepping/ramping.
Stepping and auto step/ramp limits are defined by the EZ-Checkä HI and LO settings. The step direction changes when a limit is

reached.
Figure 1 will show how the Step/Ramp Parameters are used to configure automatic stepping/ramping.

Figure 1

0

4

8

12

16

20

24

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

time

mA

Note: The Model 535’s ability to detect overload/undervoltage conditions may be limited by the rate of change in the output when

using automatic stepping/ramping. Turn auto step/ramp off while connecting or disconnecting the Model 535.

Quick Reference Bar Graph

The Quick Reference Bar Graph indicates the input and output level on the Model 535 in % of 4-20/10-50 mA with 1% resolution.

ramp

time

soak time

step size = (EZ-Step HI – EZ-Step LO)

¸

# of steps

step time = ramp time

¸

# of

t