Specifications, General specifications, Model 535 operating instructions – Practical Instrument Electronics 535 User Manual

Page 6

Advertising
background image

Model 535 Operating Instructions

82 E. Main Street Suite 3.14 · Webster, NY 14580

Tel: 585-872-9350 · Fax: 585-872-2638

[email protected] · http://www.piecal.com/

Practical Instrument Electronics, Inc. Copyright ã 2011. All rights reserved.

535-9002 Rev G 7/13/11

6-8

2 Wire Transmitter Simulate

Substitute the Model535 for a 2 wire transmitter. Use

the EZ-Dialä Knob and EZ-Checkä Switch to adjust loop

current. At least 2 volts of loop power is required, else

the display flashes “CHECK LOOP SUPPLY.”

Source Volts

The Model 532 sources 0.000-24.000 volts. This is useful

for powering transmitters and receiver equipment. Use

the EZ-Dialä Knob and EZ-Checkä Switch to adjust

output voltage. The display flashes “LOW W” when the

output is overloaded.

Specifications

General Specifications:
Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration with >1 MW load

on external banana jack access to 10 W current sense (Model 535A).

Operating Temperature Range

-20 to 60 °C (-5 to 140 °F)

Storage Temperature Range

-30 to 60 °C (-22 to 140 °F)

Relative Humidity Range

10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing

Size

7.00 X 3.30 X 2.21 inches (177.8 x 83.8 x 56.1mm)

Weight

14.0 oz (397 grams)

Battery

4 - AA Alkaline Optional 120 VAC 50/60 Hz AC adaptor included

Miscellaneous

Low battery indication with nominal 1 hour of operation left
Over-voltage protection to 120 Vrms (rated for 30 seconds) or 240 Vrms (rated for 15

seconds)
Bar graph display with 1% resolution of 4-20/10-50 mA signal scale
High contrast graphic liquid crystal display with 0.45” (11.4 mm) high digits

Advertising