MEASURE WITH MAM SENSOR

The MAM sensor is ready to measure dose and HVL for all

anode/filter combinations without changing any settings.

Place the connected sensor centered in the field, with the

crosshair 6 cm (4 cm in the US) from the examination table

front edge (A). The angle in the horizontal plane has

neglectable impact on the measurement result.
For kVp measurements, select your anode/filter combination

in the sensor settings, or by swiping right from the home

screen to access the quick settings.

Note!

When measuring kV on Mo/Rh, use the included 2 mm Al filter. Follow the instructions printed on

the filter.

Expose.

Read the result.

RaySafe X2 User Manual

26