Setting: stop delay – RaySafe X2 Manual User Manual

Page 33

Advertising
background image

SETTING: STOP DELAY

The stop delay setting defines for how long the instrument will wait for more radiation to include in the same

measurement.
Use a longer stop delay when measuring on a pulsed fluoroscopy system or to get a prepulse recorded into

the same measurement as the following ordinary exposure.

Stop delay setting longer than the time between pulses (t>T), will result in one long measurement. Note that

you have to wait as long as the stop delay setting (t) after the last exposure before the measurement ends

and values will be displayed.

Stop delay setting shorter than the time between pulses (t<T), will result in many short measurements (one

for each pulse).

RaySafe X2 User Manual

33