RaySafe i2 Installation and maintenance manual – measurement performance verification

meASURement PeRfoRmAnCe veRIfICAtIon

the purpose of this test is to verify the measurement performance of dosimeters using scattered radiation from
diagnostic X-ray units. Unfors RaySafe recommends the following arrangement:

CAUTION:

Do not expose yourself to X-ray radiation.

you need the following equipment to perform a measurement performance verification:

• A reference instrument measuring H

p

(10) (for example a known reference dosimeter, or some other meter).

• two 20x20x5 cm

3

PmmA phantoms or similarly scattering material.

• dosimeters for test

• Real time display

follow the instructions below:

1. Place one of the phantoms on the patient support.

2. Place the other phantom at the same height, but 0,5 m on the side from the first one.

3. Place the additional H

p

(10) meter on the center of the second phantom.

4. Place the dosimeters that have to be tested on the phantom around the additional H

p

(10) meter.

5. expose with a field covering the first phantom, at 80 kvp set, 4 mA, for 60s. If needed, adjust the distance

or the mA to get a dose rate of about 1 – 10 mSv/h.

6. Compare the dosimeter readings with the additional H

p

(10) meter. Readings should be within specifications

for the different meters.