Pressing force – Retsch PP 40 User Manual

Page 20

Advertising
background image

08.02.2012

Retsch GmbH

20

0002

Pressing force

5 – 40 T

The PP40 is started with the preselected pressing time and the preselected pressing force.

For press ring ø40/35x14 the pressing force is limited to 20
t. (A higher pressing force can cause a deformation of the
ring and a destruction of the pellet).


Fig.12

The PP40 is started with the preselected pressing force and pressing time.

Programme mode

Take over parameters

Here all the previously set parameters, such as pressing
force, pressing time, build-up time, holding time,
decreasing time, can be stored in a store.

Set the required parameter.

Change to „Programme mode“ in the menu, press the setting knob and confirm „Take over
parameters” again. The menu shows „Pressing programme“ and on the right the storage
location number.

Turn the setting knob to the right until you have found a pressing programme with empty
storage locations, or one which you wish to overwrite.

By pressing the setting knob you can reserve the selected storage location.

You can now choose between „Store parameter?” or “Cancel”.

You will then be returned again into the “Programme mode” level.

Programme mode

Change programme

Here all previously stored parameters, such as pressing
force, build-up time, holding time, deceasing time , can be
changed. It is also possible to enter new parameters.

Select „Programme mode“, „Change programme“ and confirm with the setting knob. The
display shows the pressing programme with storage location number again.

To select the „Pressing programme“ to be changed press the setting knob; only the storage
location numbers are inverse – the pressing programme is changed by turning the setting
knob.

Confirm the milling programme to be changed by pressing; you can now change the
parameters.

After this you can „Store (the changed) parameters” or “Cancel”.

You will then be returned again to the “Programme mode” level.

Programme mode

Delete programme

Here all the previously stored parameters can be deleted.

Select „Programme mode“, press the setting knob and confirm „Delete programme” again.

Select the pressing programme to be deleted by turning the setting knob, and confirm by

Advertising