SciGene Hybex Microsample Incubator User Manual

Page 11

Advertising
background image

Hybex® Microsample Incubator — USER MANUAL

Version 1.2H, October 2014 10 (408) 733-7337, [email protected]

IV. MAINTAINING YOUR HYBEX SYSTEM

A. Powering Off

Turn the power switch to the OFF position and unplug the power cord before performing
any service procedure.

B. Cleaning the Instrument

Clean outer metal surfaces and the heated lid using a soft cloth and mild, detergent-based
cleaner. Avoid abrasive cleaners that can scratch surfaces. Do NOT use caustic or strongly
alkaline solutions (e.g., strong soaps, ammonia, or bleach).

If running radioactive or biohazardous reactions, consult your institution’s radiation safety
officer or biosafety office for assistance.

Turn the power switch to the OFF position and unplug
the power cord before performing any service procedure!

Tournez le commutateur d'alimentation sur la position OFF et débranchez
le cordon d'alimentation avant d'effectuer toute opération de service!

Remove liquid or condensation (wipe dry) while instrument
is OFF and block is cool or at room temperature!

Enlever le liquide ou la condensation (essuyer), tandis que l'instrument
est éteint et le bloc est frais ou à température ambiante!

C. Checking and Replacing Fuses

There are two fuses located in a removable fuse block below the power cord receptacle on
the back of the unit. A blown fuse appears dark. Always replace fuses with those of the
same amperage and voltage as shown on the label below the fuse block.

Turn the power switch to the OFF position and unplug
the power cord before performing any service procedure.

To remove fuses:

1. Insert a small, flat blade screwdriver

into the tab recess just below the plug
receptacle.

2. Push down to release the fuse block.
3. Slide fuse holder out from the fuse

block while holding the retaining tab
out of the way.

4. Gently pry out the fuses.

Advertising