XI. SERVICE

A. Lockout Procedure

If a malfunction or need for service occurs, comply with the following lockout
procedure for restricting use of the instrument. All users, upon observing that the
instrument is locked out for servicing or maintenance shall not attempt to start,
energize, or use the equipment.

1. Notify all affected users that servicing or maintenance is required and that the

instrument must be shut down and locked out to perform the servicing or
maintenance.

2. If the machine is operating, shut it down by the normal stopping procedure (turn

off baths and main power or press the EMO switch).

3. Disconnect the power cord from the wall socket and from the instrument so

that the machine is isolated from energy sources (Fig. 51).

4. Twist and remove the fuse (Fig. 52). Place in a locked safe.

5. Contact your SciGene representative to arrange for service.

Fig. 51. Disconnect power cord.

Fig. 52. Remove fuse.

B. Fuse Replacement

The fuse may be replaced by turning off and unplugging the unit, then gently twisting
and removing the old fuse. Replace with a new fuse (slow blow) with the same
electrical rating (250V, 20A

for 115V or 250V, 10A for 230V)

.