IV. MAINTAINING YOUR OVEN

A. Servicing

Turn the power switch to the OFF position and unplug the power cord
before performing any service procedure.

1. Checking and Replacing Fuses

Fuses are located in a removable fuse block below the power cord
receptacle on the back of the unit.

Unplug the cord and using a flat blade screwdriver, remove the
fuse block as shown in the illustration. Remove the fuse holder
from the block and gently remove the two fuses. A blown fuse will
appear dark.

Always replace a blown fuse with fuses matching the amperage
and voltage as shown on the label below the fuse block.






















 

  

















 











 

 





Digital Thermometer Jack

Use small screwdriver behind
tab to open Fuse Block

Fuse

Block

Back of Unit

Push clip up to

release Fuse Holder