SciGene 777 Microarray Oven User Manual

Page 12

Advertising
background image

Model 777 Microarray Oven — USER MANUAL

Version 2.1, February 2011 (408) 733-7337, [email protected]

11

V. TROUBLESHOOTING

Symptom

Cause

Solution

Rotator does not turn
after closing door

Rotator control is in the off
position

Turn on rotator

Door does not shut or
is difficult to open

Latch is improperly
adjusted

Adjust threaded
rod on inside of
latch

Temperature is erratic

Thermal controller is
defective

Replace controller

VI. SPECIFICATIONS

Electrical (Oven Unit)
Cat. #1077-00-1

115V AC; 50/60 Hz; 350 W

Cat. #1077-00-2

230V AC; 50/60 Hz; 350 W

Dimensions
Interior Chamber

14 W x 14 D x 14 H inches
36 W x 36 D x 36 H cm

Exterior

18 W x 17 D x 22 H inches
45 W x 40 D x 56 H cm

Weight
Net

70 lbs (32kg)

Gross

75 lbs (34kg) Includes shipping carton

Performance and Controls
Temperature Range

Ambient + 5°C to 99°C

Temperature Control

± 0.1°C

Heat up Time

3°C per minute

Temperature Controller

Digital PID, single loop

Temperature Display

Actual / Set Dual LED

Rotation Speed

Up to 80 RPM

Chamber

Stainless Steel

Digital Thermometer Output Thermocouple

VII. ROTATORS AND ACCESSORIES

Cat. #

Description

UOM

1070-20-0

Rotator for 20 Agilent SureHyb and Original Version
Hybridization Chambers.

Each

1077-10-0

Rotator for 48 Affymetrix GeneChips.

Each

1077-30-0

Rotator for 96 Affymetrix GeneChips.

Each

1051-52-0

Digital thermometer with NIST certificate of calibration

Each

Advertising